【SUMMARY】
Opioid analgesics comprise a large group of substances which control pain by depressing the central nervous system. Large doses of Morphine can produce higher tolerance levels and physiological dependency in users, and may lead to substance abuse. Morphine is excreted unmetabolized, and is also the major metabolic product of codeine and heroin. Morphine is detectable in the urine for several days after an opiate dose.
【DIRECTIONS
FOR CASSETTE USE】
Allow the test, urine specimen, and/or controls to reach room temperature
(15-30°C) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test
cassette from the sealed pouch and use it within one hour.
2. Place the test cassette on a clean and level surface. Hold the dropper
vertically and transfer 3 full drops of urine (approx. 120μL) to the
specimen well (S) of the test cassette, and then start the timer. Avoid
trapping air bubbles in the specimen well (S). See the illustration below.
3. Wait for the colored line(s) to appear. Read results at 5 minutes. Do not
interpret the result after 10minutes.
【DIRECTIONS
FOR PANEL USE】
Allow the test, urine specimen, and/or controls to reach room temperature
(15-30ºC) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test panel
from the sealed pouch and use it within one hour.
2. Remove the cap.
3. With the arrow pointing toward the urine specimen, immerse the test panel
vertically in the urine specimen for at least 10 to 15 seconds. Immerse the
strip to at least the level of the wavy lines, but not above the arrow on the
test panel.
4. Replace the cap and place the test panel on a non-absorbent flat surface.
5. Start the timer and wait for the colored line(s) to appear.
6. The result should be read at 5 minutes. Do not interpret the result after 10
minutes.
【DIRECTIONS
FOR STRIP USE】
Allow the
test, urine specimen, and/or controls to reach room temperature (15-30ºC) prior
to testing.
1. Bring the pouch to room temperature before opening it. Remove the test Strip
from the sealed pouch and use it within one hour.
2. With arrows pointing toward the urine specimen, immerse the test Strip
vertically in the urine specimen for at least 10-15 seconds. Do not pass the
maximum line (MAX) on the Test Strip when immersing the strip. See the
illustration below.
3. Place the test Strip on a non-absorbent flat surface, start the timer and
wait for the colored line(s) to appear. Read results at 5 minutes. Do not
interpret the result after 10 minutes.