【SUMMARY】
Benzodiazepines are medications that are frequently prescribed for the symptomatic treatment of anxiety and sleep disorders. They produce their effects via specific receptors involving a neurochemical called gamma aminobutyric acid (GABA). Because they are safer and more effective, Benzodiazepines have replaced Barbiturates in the treatment of both anxiety and insomnia. Benzodiazepines are also used as sedatives before some surgical and medical procedures, and for the treatment of seizure disorders and alcohol withdrawal.
Risk of physical dependence increases if Benzodiazepines are taken regularly (e.g., daily) for more than a few months, especially at higher than normal doses. Stopping abruptly can bring on such symptoms as trouble sleeping, gastrointestinal upset, feeling unwell, loss of appetite, sweating, trembling, weakness, anxiety and changes in perception. Only trace amounts (less than 1%) of most Benzodiazepines are excreted unaltered in the urine; most of the concentration in urine is conjugated drug. The detection period for the Benzodiazepines in the urine is 3-7 days.
【DIRECTIONS
FOR CASSETTE USE】
Allow the
test, urine specimen, and/or controls to reach room temperature (15-30°C) prior
to testing.
1. Bring the pouch to room temperature before opening it. Remove the test
cassette from the sealed pouch and use it within one hour.
2. Place the test cassette on a clean and level surface. Hold the dropper
vertically and transfer 3 full drops of urine (approx. 120μL) to the specimen well (S) of the
test cassette, and then start the timer. Avoid trapping air bubbles in the
specimen well (S). See the illustration below.
3. Wait for the colored line(s) to appear. Read results at 5 minutes. Do not
interpret the result after 10minutes.
【DIRECTIONS
FOR PANEL USE】
Allow the
test, urine specimen, and/or controls to reach room temperature (15-30ºC) prior
to testing.
1. Bring the pouch to room temperature before opening it. Remove the test panel
from the sealed pouch and use it within one hour.
2. Remove the cap.
3. With the arrow pointing toward the urine specimen, immerse the test panel
vertically in the urine specimen for at least 10 to 15 seconds. Immerse the
strip to at least the level of the wavy lines, but not above the arrow on the
test panel.
4. Replace the cap and place the test panel on a non-absorbent flat surface.
5. Start the timer and wait for the colored line(s) to appear.
6. The result should be read at 5 minutes. Do not interpret the result after 10
minutes.
【DIRECTIONS
FOR STRIP USE】
Allow the test, urine specimen, and/or controls to reach room temperature
(15-30ºC) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test Strip
from the sealed pouch and use it within one hour.
2. With arrows pointing toward the urine specimen, immerse the test Strip
vertically in the urine specimen for at least 10-15 seconds. Do not pass the
maximum line (MAX) on the Test Strip when immersing the strip. See the
illustration below.
3. Place the test Strip on a non-absorbent flat surface, start the timer and
wait for the colored line(s) to appear. Read results at 5 minutes. Do not
interpret the result after 10 minutes.